Operating instructionsWasher-disinfectorG 7804To avoid the risk of accidents ordamage to the machine it isessential to read these instructionsbefore i
~When using cleaning agents andspecialised products, it is essential thatthe manufacturer's instructions arefollowed. Only use such chemicals for
~Do not sit or lean on the openeddoor. This could cause the machine totip up and be damaged or cause aninjury.~Be careful when sorting items withsharp
Accessories~Only use genuine Miele spare partsand accessories with this machine. Theuse of non-Miele parts and accessorieswill invalidate the guarante
Disposal of the packingmaterialThe transport and protective packinghas been selected from materials whichare environmentally friendly for disposaland
Electro-thermal door lockThe machine is equipped with anelectro-thermal door lock.The door can only be opened when:^the electricity supply to the mach
To open the door with theemergency releaseThe emergency release should onlybe used when the door cannot beopened normally, e.g. in the event ofa power
The water needs to be softened toavoid calcium deposits building up oninstruments and utensils and in themachine itself.To ensure a steady supply of s
Settings°d(Germanscale)mmol/l °f Setting12345678910111213141516171819 *)20212223242526272829303132333435363738394041424344450.20.40.50.70.91.11.31.41.
Filling the salt reservoirPlease only use special coarse grainedreactivation salt with granules ofapprox. 1-4 mm. Do not use other typesof salt, e.g.
Salt indicatorWhen the salt reservoir is empty the kindicator light b comes on to remindyou to fill the reservoir.Reactivation takes place automatical
Description of the machine. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Guide to the machine . . . . . . . . .
The duo dispenser incorporated in thedoor has a capacity of approximately2 x 180 ml. for liquid agents.Important:Both reservoirs must always be filled
^Add neutralising agent until the levelindicator is dark (see arrow onillustration). Use a funnel ifnecessary.^ Screw the caps back on and mop upany s
Setting the dosageThe dosage adjuster in both openingshas settings from 1-6 (1-6 ml). They areset to "5" (10 ml) at the factory as therecomm
As a chargeable optional extra thismachine can also be fitted with a“dispenser pump for liquid cleaningagent” (DOS-Module G 10). This isconnected exte
,Only use cleaning agentssuitable for washer disinfectors. Donot use detergents formulated fordomestic dishwashers.Liquid cleaning agents (mildlyalkal
Taking the effect on the environmentinto consideration, please bear thefollowing points in mind when selectinga cleaning agent:^What level of alkalini
This washer-disinfector normallyrequires a top and bottom basket or amobile (injector) unit.Various inserts and special baskets arealso available for
Note^Load items so that water will comeinto contact with all surfaces. Thisensures that they will be properlycleaned.^Do not place items to be cleaned
Loading examples:E 350 Mobile unit "Intermiel"For narrow necked glassware,complete with 15 jets 4x1 mm/160 mmhigh, 18 jets 6x1 mm/220 mm hig
U 874 Bottom basket/carrier (empty)For various insertsE 109Stainless steel half insert for 21 beakersup to 250 cm3, Erlenmeyer flasks,conical flasks e
Programming special functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 341. Rinse 2 ...
Adjusting the top basketThe top basket can be adjusted aboveand below the middle position by 2 cm.Depending on the position of the topbasket, the bask
Switching on^Close the door.^Open the stopcocks (if turned off).^Press the I-0 button.When the programme selector is at theSTOP position f a point lig
Changing a programmeProvided the Start button has not yetbeen pressed, a wrongly selectedprogramme may be changed asdescribed below (if it has beenpre
Switching off^Press and release the I-0 button.Note:As the machine is fitted with an electricdoor lock, the door can only be openedif the I-0 button i
For future servicing work, pleasedocument any alterations to the stan-dard factory settings.Enter any alterations in the fieldValue set:_________.Gene
2. Altering the temperatureand/or the temperature holdingtime in the “Main wash” and“Final rinse” programmestagesPlease make a note of any changesto t
"Final rinse" temperature:^Turn the programme selector to theSTOP position f (the programminglevel is only displayed in this switchposition)
3. Venting and setting thedosage on "DOS modules G 60and C 60" (optional extra)^Set the dosage selector on theDOS-Module to position 10 (app
4. Delay start and setting thedelay start timeThe start of a wash programme can bedelayed by up to 24 hours in 30 minutestages.If you set a very long
5. Audible tone (Buzzer)This special function allows you toprogramme a buzzer (lasting up to amaximum of 30 seconds) for thefollowing functions:–Progr
The G 7804 washer disinfector is a frontloading machine designed specificallyfor processing laboratory utensils andglassware.In general these include
6. Re-setting all parameterchanges back to the factorysettings.^Turn the programme selector to theSTOP position f.^Switch off the machine.^Press and h
Cleaning the filters in the washcabinet,This machine must not be usedwithout all the filters in place.The filter combination in the base ofthe wash ca
Cleaning the fine, flat and micro-finefilters^Remove the coarse filter.^ Remove the fine filter (if fitted) frombetween the flat and the micro-finefil
Cleaning the spray armsWaste particles can lodge in the sprayarm jets.It is therefore important to inspect thespray arms regularly.^Use a sharp pointe
Cleaning the drain pump andnon-return valveIf water has not been pumped away atthe end of a programme the drainpump or the non-return valve might bebl
Cleaning the filters in the waterinletFilters are incorporated in the screwconnection of the water inlet hose toprotect the water inlet valve. If thes
With the aid of the following explanatorynotes, minor faults can be correctedwithout contacting the Miele ServiceDept.Any work on the electrical compo
After Sales ServiceWhen a fault first occurs, pleasecheck that the fault has not beencaused by incorrect operation. Thiswill help avoid an unnecessary
Please refer to the installationdiagram supplied with the machine.,Furniture and fittings installednear the machine must be of acommercial standard ab
Position and secure the machineTo ensure stability the machine mustfirst be aligned and then screwed to theworktop.^Open the door. Secure the machinet
a On/Off button (I-0)b Door releasec Displayd Delay start indicator lighte Drying indicator lightf Programme start indicator lightg Start buttonh Dryi
,All electrical work must becarried out by a suitably qualifiedand competent person inaccordance with local and nationalsafety regulations (BS 7671 in
To remove the service panel andplinth:,Disconnect the machine from themains electricity supply.^ Remove fixing screws a from theservice panel.^ Take h
Connection to the water inlet,Water in the machine is notsuitable for drinking.–The machine must be connected tothe water supply in accordance withloc
,The inlet hose must not beshortened or damaged in any way(see illustration).See also the installation diagramsupplied.AD-water connection (pressurise
Drainage–The drainage system is fitted with anon-return valve which prevents dirtywater from flowing back into themachine via the drain hose.–The mach
Height: 85 (82) cmWidth: 60 cmDepth: 60 cmVoltage: see data plateConnected load: see data plateFuse rating: see data plateMains connection cable: appr
Programme When to use Cleaning agent(when not using liquidcleaning agent)Follow manufacturer’sinstructionsASHORTFor laboratory utensils and glassware
Programme sequencePre-wash Mainwash1)Rinse 1 Rinse 22)Finalrinse1)Drying(Additionalfunction)XDOS 170°C/5’(X)DOS 3(AD)X70°C/3’AD(DOS 2)(X)XDOS 170°C/5’
58
a Connections for DOS moduleOptional dispenser pump forliquid cleaning agent and forneutralising agent- at the back of the machine -b Filter combinati
M.-Nr. 05 841 051 / 01Alteration rights reserved / 3108
This machine complies with allrelevant local and national safetyrequirements. Incorrect use can,however, lead to personal injury anddamage to property
~Please follow the advice oninstallation in these instructions and theseparate Installation drawing.~Before connecting the machine tothe mains supply,
~During installation, maintenance andrepair work, the machine must bedisconnected from the mains electricitysupply. It is only completely isolatedfrom